Open original FDA record ↗Recall record
- FDA identifier
- Z-1042-2015
- Company in FDA record
- Zimmer, Inc.
- FDA classification date
- Feb 3, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for fracture of the tip of the product. The fractured tip is approximately 4mm x 7mm x 3.5mm in size. Retained tip may result in: pain, soft tissue irritation, soft tissue damage, or increased wear due to contact with the fractured tip; Revision surgery to remove the fractured tip
Source facts
- Action
- On 11/25/2014, URGENT MEDICAL DEVICE RECALL LOT SPECIFIC notifications were sent to the affected distributors, hospital Risk Managers and Surgeons with instructions for returning the affected product. All distributors were notified via electronic mail, and all hospital Risk Managers, Surgeons, and distributors with affected inventory were also notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call the customer call center at 1-800-348-2759.
- Initiated
- 2014-11-25
- Root cause
- Process control
- Terminated
- 2015-12-01
- Distribution
- Worldwide distribution: US nationwide, including Puerto Rico, South Korea, Singapore, Dominican Republic, Canada, Japan, and Germany.
- Recall number
- Z-1042-2015
- Recall status
- Terminated
- CDRH record ID
- 132553
- FDA classified
- 2015-02-03
- Recall event number
- 70105