Z-1042-2024: The Prestige Coil System is a product family of embolic coils with associated delivery system components.

Recall record Z-1042-2024 for BALT USA, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1042-2024
Company in FDA record
BALT USA, LLC
FDA classification date
Feb 8, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Discoloration was identified along the delivery pusher which was attributed to corrosion of the hypotube. The material which caused the corrosion was identified as remnants of solder flux residue remaining on the hypotube component of the delivery pusher. Due to the location of the delivery pusher within the delivery system, it is possible for the discoloration to mechanically break off from the delivery pusher, leaving the potential for the material to flow through the delivery system (i.e. microcatheter), and into the patient vasculature causing foreign emboli.

Source facts

Action
On 01/03/2024, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter via priority mail to customers informing them that Balt USA has became aware of a situation where discoloration along the delivery system pusher component of the Optima Coil System was identified during in-house inspection of released finished good products. The discoloration was observed on the outside of the PET jacket along the hypotube component. Customers are instructed to: 1) Stop or cease use of the affected recall product. 2) Inform their hospital staff, such as safety officers, pharmacists, purchasing agent, head of neuroradiology and the neuroradiology staff, physicians including but not limited to endovascular neurosurgeons and interventional neurologists, as well as any other person if deemed necessary of the recall. 3) Identify and locate the Optima Coil System products concerned by this recall procedure. 4) Collect and put in quarantine the Optima Coil System products concerned by this recall procedure, then return them to Balt USA through the usual RGA (Return Goods Authorization) procedure by contacting Balt USA s Customer Service department. 5) Keep informed by Balt USA about the status of every unit of Optima Coil System products concerned by this recall procedure. For questions or additional assistance, contact Balt USA Quality Department at: Contact: Quality Department email: QA@baltgroup.com Balt USA, LLC 29 Parker | Irvine CA, 92618 | USA Telephone: +1 949.788.1443 | Fax: +1 949.788.1444
Initiated
2024-01-03
Root cause
Material/Component Contamination
Distribution
U.S.: AZ, CA, CO, FL, GA, HI, IL, KS, MA, ME, MO, MS, NC, NE, NH, NJ, NY, OH, OK, PA, SC, SD, TN, TX, VA, VT, WI and Puerto Rico O.U.S.: Argentina, Australia, Brazil, Canada, Chile, Columbia, Dominican Republic, Ecuador, France, Indonesia, India, Japan, Korea, Malaysia, Mexico, Panama, Paraguay, Peru, Russia, Uruguay, and Vietnam.
Recall number
Z-1042-2024
Recall status
Open, Classified
CDRH record ID
205405
FDA classified
2024-02-08
Recall event number
93751

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗