Open original FDA record ↗Recall record
- FDA identifier
- Z-1043-2016
- Company in FDA record
- Aesculap, Inc.
- FDA record creation date
- Mar 3, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Aesculap Inc. US has initiated a recall on Tibial and Femur extension sterile pressfit stem implant packaging that are used for Columbus Revision/Enduro Knee Implants because it is labeled as Cementless. US product is indicated in the US for use with Bone Cement only.
Source facts
- Action
- Aesculap sent an Urgent Medical Device Correction letter to Distributors and Sales Rep. The letter identified the affected product, problem and actions to be taken. Customers were notified to add the over label to product packaging. Telephone conferences with Distributors and Sales Rep reiterated need for representative to inform Surgical Handler of the label when providing product for review. Customers were instructed to complete the attached Product Correction Acknowledgement Form and return to Aesculap Quality Assurance department by faxing the form to 610-791-6882 or e-mail to val.strawn@aesculap.com, two (2) weeks of receipt, even if the total inventory in your possession is zero (0).
- Initiated
- 2015-11-30
- Root cause
- Packaging change control
- Terminated
- 2016-10-13
- Distribution
- US Nationwide Distribution
- Recall number
- Z-1043-2016
- Recall status
- Terminated
- CDRH record ID
- 142358
- FDA record created
- 2016-03-03
- Recall event number
- 72853