Z-1043-2017: Optilite Freelite Lambda Free Kit Product Code: LK018.OPT, LK018.OPT.A Protein testing kit.

Recall record Z-1043-2017 for The Binding Site Group, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1043-2017
Company in FDA record
The Binding Site Group, Ltd.
FDA classification date
Jan 18, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Batch numbers listed are reporting Lambda Free Light Chain results >114mg/L that are on average 33% lower than other Optilite Lambda Freelite batches. Therefore users moving to this batch will see results >114mg/L giving lower results and conversely when switching to the next batch results will be seen to shift upwards.

Source facts

Action
On 07/05/16 notification letters were sent to customers. The Binding Site Group requests the following: - Immediately Quarantine & Dispose of any LK018.OPT lot 405566 kits within stock; - Remove any affected lots from the Optilite instrument; - Patients with Lambda Freelite results >114mg/L should be reviewed in consultation with clinicians, alongside other clinical findings and retested where applicable; - Notify local Binding Site representative to request replacement kits; - Return completed E-Back Form to local Binding Site Representative; The firm recommended that the notice be circulated to those within the organization who need to be aware.
Initiated
2016-07-05
Root cause
Under Investigation by firm
Terminated
2017-07-03
Distribution
U.S. distribution to the following; AR, PA, GA, NC, KY, VA, CA. Foreign distribution to the following; Uk, Germany, France, Italy, Spain, Czech Republic, Portugal, Belgium.
Recall number
Z-1043-2017
Recall status
Terminated
CDRH record ID
152050
FDA classified
2017-01-18
Recall event number
76048

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗