Open original FDA record ↗Recall record
- FDA identifier
- Z-1043-2019
- Company in FDA record
- Smith & Nephew, Inc.
- FDA record creation date
- Mar 22, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Higher than anticipated occurrence of bone fracture during the use of the XR Tibia Posterior Punch
Source facts
- Action
- The firm, smith&nephew, initiated the recall by email with an "Urgent Medical Device Recall Notice" letter dated 12/10/18 to the customers on 12/10/2018. Letter describes the product, problem and actions to be taken. The customers were instructed to do the following: -Inspect your inventory and locate any devices from the listed product and batch numbers and quarantine them immediately. - If you are a distributor, you must notify your customers of the field action and ensure that these actions are carried out. -Complete and return the Response Form to FieldActions@smith-nephew.com or fax to 901-566-7975. Please Note even if you have no product to return, this form must be completed, signed and returned. If you have any questions or concerns regarding this recall please contact FieldActions@smith-nephew.com.
- Initiated
- 2018-12-10
- Root cause
- Device Design
- Terminated
- 2020-11-03
- Distribution
- US Distribution to OR and TN; and Internationally to: Japan.
- Recall number
- Z-1043-2019
- Recall status
- Terminated
- CDRH record ID
- 169782
- FDA record created
- 2019-03-22
- Recall event number
- 81766