Magnet Quench, unintended- magnet of a mobile MRI system quenched into the exam room.
Source facts
Action
The firm issued an Update Instruction package to consignees 12/07. As an immediate measure, the recalling firm is initiating field checks of mobile MRI systems per Update Instruction MR039/07/S. This update defines a safety inspection for mobile systems, including checks for potential ice formation and proper configuration of venting assemblies.
Initiated
2007-12-01
Root cause
Device Design
Terminated
2012-10-01
Distribution
Nationwide Distribution
Recall number
Z-1044-2008
Recall status
Terminated
CDRH record ID
67974
FDA classified
2008-03-25
Recall event number
46245
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.