Reason stated in the source
Factor VIII quality control results are being recovered below the assigned ranges for STA System Control N+P.
Source facts
- Action
- Stago sent a recall letters via email dated August 23, 2011 to all affected customers . The letter identified the product, description of problem, and corrective actions to be taken. Customers were asked to immediately examine their inventory and follow the corrective actions provided. Customers were instructed to complete the response form attached to the letter. Because of incorrect manufacturer's instructions included in the initial letter another letter was emailed by the firm on August 30, 2011. For questions or technical assistance for updating the parameters please call the Diagnostica Stago 24 hour hotline at 1-800-725-0607.
- Initiated
- 2011-08-23
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-09-04
- Distribution
- Worldwide Distribution and USA (Nationwide) and including Canada
- Recall number
- Z-1044-2012
- Recall status
- Terminated
- CDRH record ID
- 107052
- FDA classified
- 2012-02-16
- Recall event number
- 61019