An anomaly was identified whereby the bolts used to fasten the counterweight to the gantry may not, in some cases, be fully tightened to the required torque specification.
Source facts
Action
An Urgent Medical Device Correction letter was sent to all affected users with a description of the problem and user corrective action steps. Letters were sent on 12/10/2010. Users are to immediately investigate any unusual noises coming from the affected product and report them to whoever is responsible for servicing the device. For customers that do have a service contract with Varian, Varian service will perform an inspection of the counterweight bolts as part of the regular Periodic Maintenance Inspection procedure. Customers are to advice the appropriate personnel of the contents of the letter. For further clarification, customers can contact their local Varian Customer Support Services Manager.
Initiated
2010-12-10
Root cause
Process control
Terminated
2011-08-02
Distribution
Worldwide Distribution.
Recall number
Z-1045-2011
Recall status
Terminated
CDRH record ID
96782
FDA classified
2011-02-08
Recall event number
57610