Skip to content

Public FDA record

Z-1045-2023: MEDLINE UNITE FOOT & ANKLE Screw, Non-Locking, Mini T8 (2.0x8mm), REF MMSN2008

Recall record Z-1045-2023 for MEDLINE INDUSTRIES, LP - Northfield. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1045-2023
Company in FDA record
MEDLINE INDUSTRIES, LP - Northfield
FDA classification date
Feb 1, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The recall is due to observed intra-operative screw failure. When pre-drilled with 1.3mm drill bit for 2.0mm screws and 1.6mm drill bit for 2.4mm we have observed screw failure due to excessive torque.

Source facts

Action
Medline issued a MEDICAL DEVICE RECALL notice to its direct consignees on 01/05/2023 by US mail and its distributors on 01/07/2023 by email. The notice explained the problem and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected items. Quarantine all affected product. All consignment inventory will be addressed by your local Medline UNITE Sales Representative. 2. Please return the completed enclosed response form listing the quantity of affected product on hand. Even if you do not have any affected product please complete and return the form, as indicated on the response form. 3. If you have transferred possession of this product to another individual, to another department or another location in your organization you are required by FDA regulations to notify them of this recall communication. Have the customers return any affected product to you for collection and return to Medline Industries, LP. FURTHER ACTION: Medline is immediately recommending the use of our 1.6mm pre-drill for 2.0mm screws, and our 1.9mm pre-drill for the use of our 2.4mm screws and 2.7mm screws. If you have any questions, please contact 866-359-1704."
Initiated
2022-12-22
Root cause
Device Design
Distribution
US Nationwide distribution.
Recall number
Z-1045-2023
Recall status
Open, Classified
CDRH record ID
198033
FDA classified
2023-02-01
Recall event number
91491

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗