Z-1045-2025: Discovery XR656 HD. X-Ray imaging system.

Recall record Z-1045-2025 for GE Medical Systems, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1045-2025
Company in FDA record
GE Medical Systems, LLC
FDA classification date
Feb 3, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.

Source facts

Action
Firm sent an "IMPORTANT ELECTRONIC PRODUCT RADIATION WARNING" email/letter to its consignees. The email/letter described the product, the problem and the actions to be taken. The consignees were instructed to do the following: You can continue to use the device as intended for clinical use, by following the instructions below: 1.If you perform an exposure and observe an on-screen message that the AEC termination limit has been reached, please acknowledge the message. 2.Take any relevant actions to adjust techniques or patient positioning for the next exposure to avoid reaching the termination limit for that exposure, if appropriate. Please ensure that all potential users in your facility are made aware of this notification and the recommended actions. Please retain this document for your records. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Initiated
2025-01-28
Root cause
Radiation Control for Health and Safety Act
Distribution
Worldwide Distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries to: Canada, Chile, China, Colombia, Coast Rica, Croatia, Cyprus, Czech Republic, Czechia, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Gibraltar, Greece, Honduras, Hong Kong, Hungary, Iceland, Indonesia, Iraq, Ireland, Italy, Jamaica, Japan, Jordan, Kenya, Korea (Republic of) Kuwait, Libya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Republic of Serbia, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Vietnam.
Recall number
Z-1045-2025
Recall status
Open, Classified
CDRH record ID
212319
FDA classified
2025-02-03
Recall event number
96208

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗