Z-1046-04: Alenti Lift Hygiene Chair; a non-AC-powered patient lift; Arjo, Inc., 50 N. Gary Ave., Roselle, IL 60172;

Recall record Z-1046-04 for Arjo, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1046-04
Company in FDA record
Arjo, Inc.
FDA classification date
Jul 20, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

There has been an increased number of tipping incidents and subsequent patient injuries.

Source facts

Action
Safety Advice Notice letters dated 6/8/04 were sent by certified mail to the end users, informing them of the possibility of tipping if the Alenti catches on other equipment; if the resident is not positioned correctly on the chair; if the Alenti is used with an inappropriate resident; if the resident is left unattended, particularly when the Alenti is in an elevated position; and if the safety belt is not used as a positioning belt. The end users were provided with a laminated hang tag to place on the lift with the listed safety aspects and warnings, a copy of the revised Alenti Operating and Product Care Instructions dated April 2004, and an Alenti safety belt for each of their lifts. The users were requested to complete enclosed customer response form indicating that they received and understood the safety advice notice, that they received the revised manual and destroyed all previous copies of the manual, that they received a safety belt for each lift, and that they made the patient lift operators aware of the safety advice notice and that they understood its contents. Any questions were directed to the Quality Assurance Department, at 1-800-323-1245, ext. 6103. Revised hang tags including instructions to check the castors on the lift to assure that they are secured and functioning properly, as well as providing the operating instructions for the safety belts, were sent to the accounts on 7/13/04.
Initiated
2004-06-08
Root cause
Other
Terminated
2004-12-13
Distribution
Nationwide and Canada
Recall number
Z-1046-04
Recall status
Terminated
CDRH record ID
33192
FDA classified
2004-07-20
Recall event number
29137

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗