- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Immediately after battery installation the devices announce a low battery fault warning and shut themselves off, or the service indicator icon starts blinking indicating a fault condition.
Source facts
- Action
- AED software revision 1.0.20 and higher incorporate software enhancements to make the AED more forgiving for extreme variations in user technique. Firm will provide distributors with field correction notice dated 10/5/05. Customers to be notified by mid-October 2005.
- Initiated
- 2005-10-05
- Root cause
- Device Design
- Terminated
- 2012-02-29
- Distribution
- Nationwide, Puerto Rico, Argentina, Australia, Austria, Belgium, Denmark, Ecuador, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Netherlands, Pakistan, Phillipines, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Thailand, UK
- Recall number
- Z-1046-05
- Recall status
- Terminated
- CDRH record ID
- 37884
- FDA classified
- 2005-07-28
- Recall event number
- 31445