Z-1046-05: HeartSine Samaritan AED Defibrillators Models: SAM 001, SAM 002, SAM 003.

Recall record Z-1046-05 for Heartsine Technologies Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1046-05
Company in FDA record
Heartsine Technologies Inc
FDA classification date
Jul 28, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Immediately after battery installation the devices announce a low battery fault warning and shut themselves off, or the service indicator icon starts blinking indicating a fault condition.

Source facts

Action
AED software revision 1.0.20 and higher incorporate software enhancements to make the AED more forgiving for extreme variations in user technique. Firm will provide distributors with field correction notice dated 10/5/05. Customers to be notified by mid-October 2005.
Initiated
2005-10-05
Root cause
Device Design
Terminated
2012-02-29
Distribution
Nationwide, Puerto Rico, Argentina, Australia, Austria, Belgium, Denmark, Ecuador, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Netherlands, Pakistan, Phillipines, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Thailand, UK
Recall number
Z-1046-05
Recall status
Terminated
CDRH record ID
37884
FDA classified
2005-07-28
Recall event number
31445

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗