Skip to content

Public FDA record

Z-1047-2016: Large Volume Paracentesis Kit is packaged in a thermaformed tray which contains slots and snap-in holders

Recall record Z-1047-2016 for Gi Supply. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1047-2016
Company in FDA record
Gi Supply
FDA record creation date
Mar 4, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Customers have reported that the pinch clamp on the RP Bag was missing. Ascetic fluid from the abdominal cavity may splash User if bag is not clamped prior to removal.

Source facts

Action
The customers will be made aware of the defect, including pictures provided and are instructed to inspect the RP Bags in the kit prior to use. If the pinch clamp is missing, the customer is instructed to discard the RP Bag and use the sterile RP Bag replacements provided.
Initiated
2016-01-28
Root cause
Nonconforming Material/Component
Terminated
2016-12-07
Distribution
Distributed in DC and the following states CA, FL, IA, KY, MA, MO, NH, NJ, NY, OH, OR, PA, SC, and TN.
Recall number
Z-1047-2016
Recall status
Terminated
CDRH record ID
143218
FDA record created
2016-03-04
Recall event number
73151

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗