Z-1047-2025: regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Conv

Recall record Z-1047-2025 for ROi CPS LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1047-2025
Company in FDA record
ROi CPS LLC
FDA classification date
Feb 4, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

There is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit will not reach the high and low ends of the PIP and PEEP pressure ranges impacting effective ventilation of the patient..

Source facts

Action
On 12/26/2024, the recalling firm issued a notice dated 12/26/2024 via email to their sole consignee notifying them of the recall of the convenience kits. The email also contained the notice issued by the manufacturer of the affected device component. The recalling firm's notice explained the reason for recall, the health risk involved, and the actions to be taken by the consignee, which included: (1) CHECK: Please review your inventory to determine if the manufacturer item numbers and lot numbers listed in the above table are in your inventory. (2) QUARANTINE: Please quarantine the kits affected by this notice and do not further distribute to end user facilities. (3) NOTIFY: Immediately notify any customers to whom you have distributed or forwarded product affected by this medical device correction notice. Provide a copy of this notice to all customers who have received impacted product and require a response. Customer responses will be requested to determine effectiveness of this recall. (4) END USERS: End users are instructed to discontinue use of the affected kits and quarantine. Kits are to be returned to the distribution location. (5) REPLY: Reply to this notice by completing the reply form attached and emailing to lacy.stewart@roiscs.com. (6) CORRECTION: ROi CPS, LLC personnel will conduct correction activities at the distribution location to ensure removal of the recalled Mercury Medical item. The enclosed reply form is for the consignee to report they have read and understand the attached notice. The consignee is to indicate whether they have forwarded the notice to all customers who have been shipped the affected product, whether or not they have the consignee has the affected product, and that they have completed the "Quantity on Hand" portion of the reply form.
Initiated
2024-12-26
Root cause
Nonconforming Material/Component
Distribution
Distribution was made to MO. There was no government/military/foreign distribution.
Recall number
Z-1047-2025
Recall status
Open, Classified
CDRH record ID
211994
FDA classified
2025-02-04
Recall event number
96094

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗