Through investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these errors are due to the current limit set for System Check LOCI results. The errors have no impact on results, but they may cause customer inconvenience. Siemens h
Source facts
Action
Siemens initiated their recall of this product on February 27, 2013 by sending an Important Customer Notification letter to consignees by FedEx. The letter identified the product, problem, and actions to be taken by the customers.
Initiated
2013-02-27
Root cause
Under Investigation by firm
Terminated
2014-01-14
Distribution
Worldwide Distribution.
Recall number
Z-1048-2013
Recall status
Terminated
CDRH record ID
116506
FDA classified
2013-04-01
Recall event number
64570