Open original FDA record ↗Recall record
- FDA identifier
- Z-1048-2015
- Company in FDA record
- Intuitive Surgical, Inc.
- FDA classification date
- Feb 3, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The Wall chart has been updated because it was noted that Wall Chart (PN 551524-03 Rev A) was inconsistent with the User Manual Adendum (PN 550986-07_B) with regards to neceassary sterilization dry time.
Source facts
- Action
- Medical Device Recall Replacement letters were sent to all customers on January 15, 2015. The letter identified the reason for the recall, the risk to health, affected products, and actions to be taken by the customer or user and the firm. Customers were asked to discard the existing EndoWrist Stapler Documentation and replace it with the new Documentation provided by the firm. Intuitive Surgical will ship replacement instruments at no charge to customers. Then customers were asked to reunite the replacement instruments with the additional components. The firm will retrain customers on the updated instructions for use. A Clinical Sales Representative or Intuitive Surgical Customer Service at 1-800-876-1310, should be contacted if further information or support is needed concerning the recall notification.
- Initiated
- 2015-01-15
- Root cause
- Error in labeling
- Terminated
- 2015-03-03
- Distribution
- Nationwide Distribution
- Recall number
- Z-1048-2015
- Recall status
- Terminated
- CDRH record ID
- 132453
- FDA classified
- 2015-02-03
- Recall event number
- 70181