Skip to content

Public FDA record

Z-1048-2023: GORE CARDIOFORM Septal Occluder, REF: GSX0030A

Recall record Z-1048-2023 for W L Gore & Associates, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1048-2023
Company in FDA record
W L Gore & Associates, Inc.
FDA classification date
Feb 1, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Septal Occluders released with incomplete quality test documentation that assesses the release of the device from the delivery system during the index procedure. A potential failure of this quality test could potentially result could result in occluder embolization, anoxic brain injury, central venous or arterial perforation or rupture.

Source facts

Action
On 12/14/22, recall notices were mailed to customers informing them to take the following actions: 1) Identify and return any unused devices. 2) Complete and return the customer response form to to FieldActionTeam@wlgore.com 3) This notice needs to be shared with those who should be aware within your institution or to any organization where potentially affected devices have been transferred. Customers with questions can contact the firm at MPDCustomerCare@wlgore.com
Initiated
2022-12-14
Root cause
Employee error
Distribution
US Nationwide distribution in the states of NY, CA, FL, CT, TX, AR, IN, MA, MO, SD, VA.
Recall number
Z-1048-2023
Recall status
Open, Classified
CDRH record ID
197593
FDA classified
2023-02-01
Recall event number
91380

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.