Open original FDA record ↗Recall record
- FDA identifier
- Z-1049-2023
- Company in FDA record
- NovaSignal Corp.
- FDA classification date
- Feb 1, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Transcranial Doppler/ultrasound system tablet has a faulty hard drive, which will cause the system to freeze and reboot. Initially, the system will reboot to a normal operable state but eventually the system will be rendered inoperable.
Source facts
- Action
- On 12/16/22, recall notices were emailed to customers. Customers were told they would be contacted by the recalling firm to facilitate the proper disposition of affected devices. If an unexpected reboot occurs, the recalling firm recommends pausing the use of affected devices until unaffected devices arrive. Complete and return the customer response form. Customers with questions can contact Customer Operations at anchal@novasignal.com
- Initiated
- 2022-12-16
- Root cause
- Nonconforming Material/Component
- Distribution
- US Nationwide and Worldwide Distribution: AZ, TX, SC, PA, IL. OUS: CA, SE, NL
- Recall number
- Z-1049-2023
- Recall status
- Open, Classified
- CDRH record ID
- 197861
- FDA classified
- 2023-02-01
- Recall event number
- 91467