Open original FDA record ↗Recall record
- FDA identifier
- Z-1049-2025
- Company in FDA record
- MEDLINE INDUSTRIES, LP - Northfield
- FDA classification date
- Jan 29, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Certain lots Sterile Infusion Catheters which were mislabeled as having a heparin coating were either distributed or manufactured into medical convenience kits and distributed.
Source facts
- Action
- Medline Industries, LP issued one IMMEDIATE ACTION REQUIRED notice and one MEDICAL DEVICE RECALL notice to its affected consignees on 12/23/2024 via email and US Mail. The IMMEDIATE ACTION REQUIRED notice explained the labeling issue with the device, potential risk, and requested the following actions be taken: 1. affected catheters be quarantined, 2. respond with quantity on hand, 3. if the product was transferred to notify those to whom the affected catheters were transferred. Upon receipt of the response. Medline will issue corrected labeling to affix to the devices. The MEDICAL DEVICE RECALL notice, which pertains to kits manufactured with the affected catheters as a component, also explained the labeling issue with the device, potential risk, and requested the following actions be taken: 1. affected catheters be quarantined, 2. respond with quantity on hand, 3. if the product was transferred to notify those to whom the affected catheters were transferred. Upon receipt of the response. Medline will issue a label to affix to the devices which will state the affected component shall be removed and discarded from further use.
- Initiated
- 2024-12-23
- Root cause
- Process change control
- Distribution
- US: FL, KS, ND, PA, WA, WV, IL, WA
- Recall number
- Z-1049-2025
- Recall status
- Open, Classified
- CDRH record ID
- 211904
- FDA classified
- 2025-01-29
- Recall event number
- 96066