Open original FDA record ↗Recall record
- FDA identifier
- Z-1050-2008
- Company in FDA record
- Physio Control, Inc.
- FDA classification date
- Mar 29, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Inoperable Device - Potential for delay in therapy or prevention of defibrillation therapy due to corrosion of the Printed Circuit Board Assemblies (PCBA).
Source facts
- Action
- The firm contacted Customers by phone and by Urgent-Medical Device Recall letters sent via return receipt requested to their consignees on 1/8/08. The letter advised consignees that the Printed Circuit Board Assemblies (PCBA) might not meet quality requirements. This manufacturing deviation could lead to eventual PCBA corrosion that could render the device inoperative and not ready for use in an emergency under battery power. The firm recommended keeping the unit in service until it could be replaced. Representatives will visit each consignee and replace units.
- Initiated
- 2008-01-08
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-09-04
- Distribution
- Worldwide Distribution --- USA including states of OK, SC, TX, and WA, and country of Germany.
- Recall number
- Z-1050-2008
- Recall status
- Terminated
- CDRH record ID
- 67227
- FDA classified
- 2008-03-29
- Recall event number
- 46351