Open original FDA record ↗Recall record
- FDA identifier
- Z-1050-2015
- Company in FDA record
- Hospira Inc.
- FDA classification date
- Feb 4, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Plum A+ infusion system pole clamp assemblies were discovered to be cracked. If the knob on the pole clamp assembly is cracked, the infuser cannot be attached to the IV pole since the knob is used to tighten and loosen the grip of the pole clamp to the pole.
Source facts
- Action
- Hospira contacted the the sole customer on June 11, 2014, to inform them of the issue of the potential impact to their Plum A+ infusers list 20792-04-25. Hospira requested that the customer to allow Hospira Field Service Representatives to inspect and replace the impact pole clamp assemblies. On June 12, 2013 Geneva Hospital approved Hospira's request to the inspection and replacement of the impacted pole clamp assemblies. On June 16, 2013 Hospira's Field Service personnel performed the on-site, visual inspection of the pole clamp assemblies and confirmed 15 units were cracked. Hospira determined to replace all 102 pole clamp assemblies that were received by Geneva Hospital. For further questions please call (800) 441- 4100.
- Initiated
- 2014-06-11
- Root cause
- Vendor change control
- Terminated
- 2016-11-28
- Distribution
- US Distribution to New York.
- Recall number
- Z-1050-2015
- Recall status
- Terminated
- CDRH record ID
- 128430
- FDA classified
- 2015-02-04
- Recall event number
- 68672