Transcranial Doppler/ultrasound system tablet has a faulty hard drive, which will cause the system to freeze and reboot. Initially, the system will reboot to a normal operable state but eventually the system will be rendered inoperable.
Source facts
Action
On 12/16/22, recall notices were emailed to customers. Customers were told they would be contacted by the recalling firm to facilitate the proper disposition of affected devices. If an unexpected reboot occurs, the recalling firm recommends pausing the use of affected devices until unaffected devices arrive.
Complete and return the customer response form.
Customers with questions can contact Customer Operations at anchal@novasignal.com
Initiated
2022-12-16
Root cause
Nonconforming Material/Component
Distribution
US Nationwide and Worldwide Distribution: AZ, TX, SC, PA, IL. OUS: CA, SE, NL
Recall number
Z-1050-2023
Recall status
Open, Classified
CDRH record ID
197869
FDA classified
2023-02-01
Recall event number
91467