Z-1050-2025: BD Pyxis MedStation ES AUX Tower REF 343 BD Pyxis " MedStation 4000 AUX Tower REF 314 The BD Pyxis" Med

Recall record Z-1050-2025 for CareFusion 303, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1050-2025
Company in FDA record
CareFusion 303, Inc.
FDA classification date
Jan 30, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to an increase in complaints of door latch assembly failures that may cause the doors to be left in an open state potentially causing inventory discrepancies, a delay in access to medication and/or diversion of medication.

Source facts

Action
On 02/14/2023 and 01/09/2024, the firm sent an "Urgent Medical Product Advisory" letter to customers informing them that BD was made aware due to an increase in customer complaints for AUX Tower Door Latch assembly failures. The latch failure can leave the door in an open state, potentially causing inventory discrepancies (including billing), a delay in access to medication and/or diversion of medication. Customers were instructed: 1. May continue to use these devices as intended and should file a complaint via the Customer Portal at https://www.bd.com/self-service or call BD Support at 1-800-727-6102, should you experience drawer or door failures. 2. If customers experience delays in access to medication due to a drawer or door failure, ensure backup procedures (which may include but may not necessarily be limited to keys, emergency medication supplies, alternate devices in proximity, etc.) are in place and staff have all been properly trained on them. 3. Promptly complete and return the enclosed Customer Response Form to acknowledge receipt of this notification and that you understand the information provided in this letter. For Distributors: 1. Provide a copy of this notification to any customers who may have received products listed in the Affected Products table above. 2. Promptly complete and return the enclosed Customer Response Form to acknowledge receipt of this notification and that you understand the information provided in this letter. For questions - contact 1-866-583-8783 Actions Taken by BD for Door Latch Failure: 1. Revised BD Pyxis Field Service Guide to include the use of a conductive contact enhancer that will help prevent corrosion and increase connectivity to the latch thus significantly reducing door failures. 2. Updating the door latch design. 3. If Auxiliary Tower Door failure occurs due to poor connection, BD will provide service support to evaluate and resolve the issue.
Initiated
2023-02-14
Root cause
Device Design
Distribution
U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Germany and Japan
Recall number
Z-1050-2025
Recall status
Open, Classified
CDRH record ID
211798
FDA classified
2025-01-30
Recall event number
96020

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗