Open original FDA record ↗Recall record
- FDA identifier
- Z-1051-2008
- Company in FDA record
- Draeger Medical, Inc.
- FDA classification date
- Mar 18, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Heating failure- The power board which controls the incubator heater may not regulate the temperature properly resulting in a high temperature alarm and a loss of temperature control within the patient compartment.
Source facts
- Action
- The firm notified consignees on 2/1/08, by an Important Safety Notice. The letter instructed users to remove any units from use until a Draeger Service Representative can replace the power board. A Service Representative will be in contact to schedule a service visit. For questions concerning this recall please contact 215-660-2418 or Anbu.arasu@draeger.com.
- Initiated
- 2008-01-31
- Root cause
- Nonconforming Material/Component
- Terminated
- 2008-05-08
- Distribution
- Worldwide Distribution-- USA including states MO, OH, and PR, and countries of Australia, Canada, Columbia, Israel, UAE, France, China, Italy, Denmark, Switzerland, and UK.
- Recall number
- Z-1051-2008
- Recall status
- Terminated
- CDRH record ID
- 68118
- FDA classified
- 2008-03-18
- Recall event number
- 46742