Skip to content

Public FDA record

Z-1051-2015: MAQUET, SERVO Humidifier 163, Rx Only, Inmed Mfg. Sdn Bhd. Catalog No: XKC01-06-8125-8, Model No. 01-06-

Recall record Z-1051-2015 for Teleflex Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1051-2015
Company in FDA record
Teleflex Medical
FDA classification date
Feb 6, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Defects: Cracks that may occur in the center of the corrugated flex tube of the 15 mm and 22 mm ISO female connector.

Source facts

Action
Consignees were notified by an Urgent - Field Safety Notice, dated 01/12/2015. The letter identified the affected devices, a description of the problem, and actions to be taken. Customers were asked to check their inventory for any affected product, to cease use and distribution, and immediately quarantine any affected product. Customers are to complete and return the Acknowledgement form and to their local MAQUET representative, who provide a return number. A credit will be issues upon receipt of the affected product. Customer should contact their local MAQUET representative at fieldactions@maquet.com with questions or for support this issue.
Initiated
2015-01-12
Root cause
Process control
Terminated
2016-05-10
Distribution
Distributed to FL, KS, MI and WV.
Recall number
Z-1051-2015
Recall status
Terminated
CDRH record ID
132643
FDA classified
2015-02-06
Recall event number
70257

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗