BD Synapsys version 2.1 allowed the potential for a test order to be associated to an incorrect culture record.
Source facts
Action
BD notified Synapsys Customer via letter UPS on 02/21/19 identified problem and action taken. Customer sites were updated remotely with a database update on January 10, 2019. Accounts asked to complete and return the CUSTOMER RESPONSE FORM BD Synapsys. The issue will be resolved by the software update syngo CT VB10A_SP1 (Service Pack 1) as soon as the update is available. For questions contact BD Customer/Technical Support at 8006388663. For customers outside the US, contact your local BD representative or distributor.
Initiated
2019-02-21
Root cause
Software design
Terminated
2020-04-17
Distribution
Worldwide Distribution - US: Nationwide in the states of OH, IL, MO, TN and the countries of Netherlands
Recall number
Z-1051-2019
Recall status
Terminated
CDRH record ID
171162
FDA record created
2019-03-23
Recall event number
82225