Z-1051-2023: Catalog Number / Product Description 0601600 2.7F Broviac CV Catheter Repair Kit 0601610 4.2 Fr Broviac

Recall record Z-1051-2023 for Bard Peripheral Vascular Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1051-2023
Company in FDA record
Bard Peripheral Vascular Inc
FDA classification date
Feb 1, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to the adhesive provided within the catheter repair kits maybe hardened/coagulated (unusable).

Source facts

Action
On 01/04/2023, the firm sent an "URGENT MEDICAL DEVICE RECALL" Letter via FedEx or email to customers informing them that the firm has received complaints regarding hardened/coagulated adhesive associated with various Catheter Repair Kits. as a result of the hardened/coagulated adhesive, there is the potential to cause delays whilst an alternative repair kit, adhesive or replacement catheter is obtained, therefore potentially prolonging surgery or necessitating exchange. Customer are instructed to: 1. Discontinue use of product listed in Attachment A, and immediately dispose of all affected product remaining in their possession in accordance with their local facility s process. 2. Share this notification with all users within their facility network of the product to ensure they are also aware of this field action. 3. If customers purchased this product from a distributor, contact the distributor for further instructions and credit resolution. 4. Complete and return the attached Customer Response Form and return it to the BD contact noted on the form whether or not they have any of the product so that BD may acknowledge their receipt of this notification and subsequently process any applicable credit regarding impacted product. 5. Indicate on the response form the quantity of product identified at their facility and confirm that this product inventory was disposed of. For Distributors (in addition to the above): Identify all customers within their distribution network that purchased any affected product and provide a copy of the attached customer letter to all customers to advise them of this field action notification on BD s behalf. Contact their BD representative if they require assistance with this process or they may also contact the number below for further assistance. North American Regional Complaint Center Phone: 1-844-8BD- LIFE (1-844-823-5433) Say Recall when prompted M-F 8am - 5pm CT Email: productcomplaints@bd.com
Initiated
2023-01-04
Root cause
Under Investigation by firm
Distribution
Worldwide Distribution: U.S.(nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV; OUS (foreign) countries of: Australia, Belgium, Canada, Hong Kong and Japan. O.U.S.: Australia, Belgium, Canada, Hong Kong, and Japan.
Recall number
Z-1051-2023
Recall status
Open, Classified
CDRH record ID
197463
FDA classified
2023-02-01
Recall event number
91327

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗