Skip to content

Public FDA record

Z-1052-2008: Draeger GT500. GlobeTrotter Transport System. Catalog Number: MU20508, Draeger Medical, Inc., Telford, PA

Recall record Z-1052-2008 for Draeger Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1052-2008
Company in FDA record
Draeger Medical, Inc.
FDA classification date
Mar 18, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Heating failure- The power board which controls the incubator heater may not regulate the temperature properly resulting in a high temperature alarm and a loss of temperature control within the patient compartment.

Source facts

Action
The firm notified consignees on 2/1/08, by an Important Safety Notice. The letter instructed users to remove any units from use until a Draeger Service Representative can replace the power board. A Service Representative will be in contact to schedule a service visit. For questions concerning this recall please contact 215-660-2418 or Anbu.arasu@draeger.com.
Initiated
2008-01-31
Root cause
Nonconforming Material/Component
Terminated
2008-05-08
Distribution
Worldwide Distribution-- USA including states MO, OH, and PR, and countries of Australia, Canada, Columbia, Israel, UAE, France, China, Italy, Denmark, Switzerland, and UK.
Recall number
Z-1052-2008
Recall status
Terminated
CDRH record ID
68121
FDA classified
2008-03-18
Recall event number
46742

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗