Z-1052-2022: HemosIL ReadiPlasTin, Part No. 0020301400

Recall record Z-1052-2022 for Instrumentation Laboratory. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1052-2022
Company in FDA record
Instrumentation Laboratory
FDA record creation date
May 10, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Variable and out of specification QC results with HemosIL ReadiPlasTin. IL will remove all lots from the market and convert customers to an alternative prothrombin time (PT) reagent, specifically HemosIL RecombiPlasTin 2G.

Source facts

Action
On April 8, 2022, the firm notified customers via email and certified mail. Customers were informed that the firm will stop shipment of product on May 9, 2022. After receipt of the material, customers must complete the conversion to HemosIL RecombiPlasTin 2G, with support as needed from your local Werfen representative, within 90 days. Until a lot of HemosIL RecombiPlasTin 2G has been received and ready for use, customers should run quality control for HemosIL ReadiPlasTin with each new reagent vial (every lot) and at least every 8 hours to identify vials with performance issues. Any product with failed quality control or quality control results which trend high should be discarded. Once customers have completed the comversion to HemosIL RecombiPlasTin 2G, any remaining unused HemosIL ReadiPlasTin should be destroyed.
Initiated
2022-04-08
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution including in the states of AZ, CA, IA, IL, IN, KS, MD, ME, MI, MN, MO, NC, NE, NY, OH, OK, SC, TX, VA, WV.
Recall number
Z-1052-2022
Recall status
Open, Classified
CDRH record ID
193157
FDA record created
2022-05-10
Recall event number
90012

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗