Open original FDA record ↗Recall record
- FDA identifier
- Z-1052-2022
- Company in FDA record
- Instrumentation Laboratory
- FDA record creation date
- May 10, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Variable and out of specification QC results with HemosIL ReadiPlasTin. IL will remove all lots from the market and convert customers to an alternative prothrombin time (PT) reagent, specifically HemosIL RecombiPlasTin 2G.
Source facts
- Action
- On April 8, 2022, the firm notified customers via email and certified mail. Customers were informed that the firm will stop shipment of product on May 9, 2022. After receipt of the material, customers must complete the conversion to HemosIL RecombiPlasTin 2G, with support as needed from your local Werfen representative, within 90 days. Until a lot of HemosIL RecombiPlasTin 2G has been received and ready for use, customers should run quality control for HemosIL ReadiPlasTin with each new reagent vial (every lot) and at least every 8 hours to identify vials with performance issues. Any product with failed quality control or quality control results which trend high should be discarded. Once customers have completed the comversion to HemosIL RecombiPlasTin 2G, any remaining unused HemosIL ReadiPlasTin should be destroyed.
- Initiated
- 2022-04-08
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution including in the states of AZ, CA, IA, IL, IN, KS, MD, ME, MI, MN, MO, NC, NE, NY, OH, OK, SC, TX, VA, WV.
- Recall number
- Z-1052-2022
- Recall status
- Open, Classified
- CDRH record ID
- 193157
- FDA record created
- 2022-05-10
- Recall event number
- 90012