Skip to content

Public FDA record

Z-1052-2023: NuVasive MD Pulse III Multimodality System REF 1881500 Pulse Software version 369-BL.US and 369-BLN.US

Recall record Z-1052-2023 for NuVasive Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1052-2023
Company in FDA record
NuVasive Inc
FDA classification date
Feb 3, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Due to an incorrect security key, customers were unable to connect to remote monitoring during spinal surgery, neck dissections, thoracic surgeries, and upper and lower extremity procedures.

Source facts

Action
Between 07/19/2022 to 12/02/2022, affected customers were contacted and verbally informed during a software update that there was an issue related to the inability to remotely connect their system and does not allow physicians who is not physically in the operating room to be able to review live neuromonitoring data remotely and provide analysis to the surgeon in the operating room. The firm is planning on sending an "URGENT: MEDICAL DEVICE RECALL" Letter on 01/13/2023, to officially inform customers that a recall is being conducted due to a remote monitoring issue that may potentially not allow a physician who is not physically in the operating room to be able to review live neuromonitoring data remotely and provide analysis to the surgeon in the operating room. Customer Instructions: -Nuvasive, Inc. will contact a Representative at the hospitals to schedule the software key upgrade. -If the security key has been updated, no further action is necessary. For questions or assistance, contact Sr. Director, Global Quality at 858-336-3421 or email pyrigoyen@nuvasive.com
Initiated
2022-07-19
Root cause
Software design
Distribution
U.S. Nationwide distribution in the states of CA, FL, NC, OH, and TX.
Recall number
Z-1052-2023
Recall status
Open, Classified
CDRH record ID
197394
FDA classified
2023-02-03
Recall event number
91294

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.