Skip to content

Public FDA record

Z-1053-2013: LEEP REDIKIT Product No. 6060 In a Loop Electrosurgical Excision Procedure, a thin, low-voltage ele

Recall record Z-1053-2013 for Cooper Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1053-2013
Company in FDA record
Cooper Surgical, Inc.
FDA classification date
Apr 3, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

LEEP RediKit Products (for use in Loop Electrosurgical Excision Procedures) have Xylocaine (Lidocaine) HCL (2%) listed on the tray lid label but this is not included in the contents of the kit, which contained Polocaine (Mepivacaine) HCL (3%) instead (which was not listed on the tray label).

Source facts

Action
CooperSurgical sent a Recall Notification dated February 26, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. CSI intends to track and quarantine returned product in a segregated location at its manufacturing facility to ensure its separation from acceptable product and prevent its misuse. CSI intends to render used product useless by physical destruction and to rework unused product by correcting the labeling for redistribution. If you have any one of these affected lot numbers in your current inventory, please contact our Customer Service Department at 800-243-2974, or simply complete the attached form and fax to 800-262-0105. Arrangements will be made to replace product in your possession at CooperSurgical's expense. Further instruction and disposition of product will be communicated to you at that time. CooperSurgical deeply regrets the inconvenience that this problem has caused.
Initiated
2013-02-26
Root cause
Labeling mix-ups
Terminated
2014-06-12
Distribution
Nationwide Distribution including Puerto Rico
Recall number
Z-1053-2013
Recall status
Terminated
CDRH record ID
116406
FDA classified
2013-04-03
Recall event number
64528

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗