Open original FDA record ↗Recall record
- FDA identifier
- Z-1054-2015
- Company in FDA record
- Cardinal Health
- FDA classification date
- Feb 4, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A small piece of the stylet sheath has the potential to shear off upon removal of the stylet from the tracheal tube.
Source facts
- Action
- URGENT: PRODUCT RECALL letters dated December 23, 2014 were sent to all customers of record. The letters included instructions for customers to: 1) quarantine the affected product; 2) notify any customers whom you may have distributed the affected product to; 3) contact the appropriate Customer Service group to arrange for return and credit of any affected product in your inventory; and, 4) complete and return the enclosed Acknowledgement Form. Customers with questions about this recall can contact Cardinal Health Quality Systems at 800-292-9332.
- Initiated
- 2014-12-23
- Root cause
- Process control
- Terminated
- 2016-08-01
- Distribution
- US: Nationwide (AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, WV, WY) + District of Columbia + Guam + US Army Facility in South Korea
- Recall number
- Z-1054-2015
- Recall status
- Terminated
- CDRH record ID
- 132452
- FDA classified
- 2015-02-04
- Recall event number
- 70185