Z-1054-2019: Omnidiagnost Classic, Product Codes 70859, 708023, 708024, 708025 Product Usage: The Philips OmniDiagnos

Recall record Z-1054-2019 for Philips North America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1054-2019
Company in FDA record
Philips North America, LLC
FDA record creation date
Mar 23, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The locking plate may be missing from the system, which may result in unexpected stand/table tilt movement that cannot be stopped by the user.

Source facts

Action
On February 28, 2019, the firm notified customers via an Urgent Field Safety Notice letter. The letter informed customers that the affected systems would be inspected for missing locking plates. Customers were asked to place a copy of the Field Safety Notice with the documentation of the system until the system has been inspected by Philips. Philips will inspect all potentially affected systems free of charge. If a locking plate is found missing, Philips will secure the system until a locking plate is installed so that use of the system can continue. You will be contacted by your local Philips representative to schedule the inspection. Philip's action will start effective from March 2019. If you need further information or support concerning this issue, please contact your local Philips representative.
Initiated
2019-02-28
Root cause
Component design/selection
Terminated
2020-08-03
Distribution
US Nationwide domestic distribution, and worldwide international distribution.
Recall number
Z-1054-2019
Recall status
Terminated
CDRH record ID
171336
FDA record created
2019-03-23
Recall event number
82313

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗