Open original FDA record ↗Recall record
- FDA identifier
- Z-1055-2017
- Company in FDA record
- The Binding Site Group, Ltd.
- FDA classification date
- Jan 19, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
If a control ring is marked after reading, the software will not flag results that are out of the specified QC range.
Source facts
- Action
- On 10/2012 notification letter was sent to customers requesting they take the following actions: 1. If have not already installed software version Vl.0.3 then please continue to work with version V1.0.2 until the above issues are resolved. 2. If have already installed software version V1.0.3, when processing results always mark the controlprior to reading the ring. 3. If are uncertain about the validity of any assays you have performed using software version Vl.0.3 please review the controlresults obtained and compare them to the specified control range for the RID assay and; " If the control results are acceptable then your assay is valid and results will be unaffected. " If the control results are not acceptable your assay should be considered invalid and your patient results should be reviewed and repeated as appropriate. On 02/2013 a second notification was sent in which customers were informed the new software was released. Customers were asked to take the following actions; 1. Upon receipt install the new software and uninstall the previou version 2. Install following the onscreen instructions. The installation process has not changed from previous versions. The instruction guide received with the RID Plate reader is still the current version. 3. After receipt of the new software disc complete and return the Eback form to The Binding Site. For further assistance contact Technical Services at The Binding Site via the email address: technicalservices@bindingsite.co.uk.
- Initiated
- 2012-10-09
- Root cause
- Software Design Change
- Terminated
- 2017-04-04
- Distribution
- U.S. distribution to the following; MA, GA, NY, IA, CA, IL, UT, NJ, NC, MD, MN. Foreign distribution to the following; UK, Germany, Spain, France.
- Recall number
- Z-1055-2017
- Recall status
- Terminated
- CDRH record ID
- 152040
- FDA classified
- 2017-01-19
- Recall event number
- 76043