Open original FDA record ↗Recall record
- FDA identifier
- Z-1055-2021
- Company in FDA record
- Aomori Olympus Co., Ltd.
- FDA record creation date
- Feb 5, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Source facts
- Action
- On January 4, 2021, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were informed that the identified list of Endo-therapy devices may have a defective seal which may compromise the sterility of the product. Customers were asked to take the following actions: 1. Immediately assess any product you have in stock to identify endotherapy products with affected lot numbers listed in the recall letter. Cease use of product and quarantine any affected product. 2. Call your Olympus customer service representative at 1-888-524-7266 to obtain a Returned Goods Authorization so that you may return the product with no charge to you. Olympus will issue a credit or replacement to your facility for any returned product. 3. Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number Recall-0391 and provide your contact information as indicated in the portal. 4. If you have distributed these devices outside your facility, please notify your customers of this corrective action immediately. Your notification to your customers may be enhanced by including a copy of this recall notification letter.
- Initiated
- 2021-01-04
- Root cause
- Under Investigation by firm
- Terminated
- 2024-03-11
- Distribution
- US Nationwide distribution.
- Recall number
- Z-1055-2021
- Recall status
- Terminated
- CDRH record ID
- 185293
- FDA record created
- 2021-02-05
- Recall event number
- 87074