Skip to content

Public FDA record

Z-1056-2009: Zimmer Coonrad/Morrey Total Elbow, Interchangeable Ulnar Assembly, plasma sprayed, extra small, right, 3

Recall record Z-1056-2009 for Zimmer Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1056-2009
Company in FDA record
Zimmer Inc.
FDA classification date
Apr 2, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The surgical technique is inadequate for extra-small sized implants, resulting in the possibility of the provisional becoming stuck in the bone.

Source facts

Action
Consignees were notified by "Urgent Device Correction" letter dated January 30, 2009 and provided revised surgical techniques for these extra-small implants. Direct questions about this recall to Zimmer, Inc. at 574-267-6131.
Initiated
2009-01-30
Root cause
Labeling design
Terminated
2009-12-22
Distribution
Worldwide Distribution, including USA, Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Denmark, England, Finland, France, Germany, Hong Kong, Hungary, India, Italy, Japan, Jordan, Latin America, Netherlands, Russia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkey and Venezuela.
Recall number
Z-1056-2009
Recall status
Terminated
CDRH record ID
77693
FDA classified
2009-04-02
Recall event number
50914

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗