Open original FDA record ↗Recall record
- FDA identifier
- Z-1056-2019
- Company in FDA record
- Philips North America, LLC
- FDA record creation date
- Mar 23, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The locking plate may be missing from the system, which may result in unexpected stand/table tilt movement that cannot be stopped by the user.
Source facts
- Action
- On February 28, 2019, the firm notified customers via an Urgent Field Safety Notice letter. The letter informed customers that the affected systems would be inspected for missing locking plates. Customers were asked to place a copy of the Field Safety Notice with the documentation of the system until the system has been inspected by Philips. Philips will inspect all potentially affected systems free of charge. If a locking plate is found missing, Philips will secure the system until a locking plate is installed so that use of the system can continue. You will be contacted by your local Philips representative to schedule the inspection. Philip's action will start effective from March 2019. If you need further information or support concerning this issue, please contact your local Philips representative.
- Initiated
- 2019-02-28
- Root cause
- Component design/selection
- Terminated
- 2020-08-03
- Distribution
- US Nationwide domestic distribution, and worldwide international distribution.
- Recall number
- Z-1056-2019
- Recall status
- Terminated
- CDRH record ID
- 171338
- FDA record created
- 2019-03-23
- Recall event number
- 82313