Skip to content

Public FDA record

Z-1057-2017: Chord-X Chordal Sizer is designed to assist in determining the proper length of chordal replacement prost

Recall record Z-1057-2017 for On-X Life Technologies, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1057-2017
Company in FDA record
On-X Life Technologies, Inc.
FDA classification date
Jan 20, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Four issues were uncovered by the recalling firm's CAPA: 1) Burrs were found to be present on Left Housing, Right Housing and Barrel components, 2) Loose plastic particulate was found within sealed primary packages, 3) Loose hair was discovered within four sealed primary packages and 4) Assembly error, wherein Left and Right Housing components were acceptably press-fit together, was discovered within one sealed primary package. One of the Left Housing press-fit pins was not pressed into the associated hole of the Right Housing, resulting in a small gap in the housing assembly.

Source facts

Action
A letter was sent to the implanting surgerons notifyijng them of the recall and attemmpting retrieval.
Initiated
2016-12-13
Root cause
Component design/selection
Terminated
2017-04-13
Distribution
CA, WA, NE, DE, OH, ME, TX
Recall number
Z-1057-2017
Recall status
Terminated
CDRH record ID
151822
FDA classified
2017-01-20
Recall event number
75951

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗