Z-1057-2019: Randox Blood Gas Control Level 2 -Model BG5002. In-vitro Diagnostic Quality control of Blood Gas analysis

Recall record Z-1057-2019 for Randox Laboratories Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1057-2019
Company in FDA record
Randox Laboratories Ltd.
FDA record creation date
Mar 26, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An error in the assignment of the control range pH in Blood Gas Control

Source facts

Action
The firm, Randox Laboratories Ltd, sent an "Urgent Medical Device Correction " letter dated 2/13/2019 to its customer on 2/13/2019 via email. The letter describes the product, problem and actions to be taken. The customer was instructed todo the following: " Please replace the amended value sheet within the kit of any remaining stock. " Discuss the contents of this notice with your Medical Director. " Randox Laboratories Ltd is not recommending a review of previous results as any issue would be observed at the time of use. " Please retain this letter with your laboratory records and forward this letter to those who may have received this product. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. If you have any questions or concerns, please contact Randox Technical Services by email at technical.services@randox.com, or telephone +44 (0)28 9442 2413 or Email: vigilance@randox.com.
Initiated
2019-02-13
Root cause
Process control
Terminated
2020-04-14
Distribution
US Distribution to state of: WV
Recall number
Z-1057-2019
Recall status
Terminated
CDRH record ID
171118
FDA record created
2019-03-26
Recall event number
82195

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗