Skip to content

Public FDA record

Z-1057-2024: AirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, without Mask, REF 2K8000,

Recall record Z-1057-2024 for Vyaire Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1057-2024
Company in FDA record
Vyaire Medical
FDA classification date
Feb 25, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.

Source facts

Action
Vyaire issued an URGENT: Customer notification Letter to its consignees on 01/10/2024 via letter delivered by UPS Next Day Air. The notice explained the problem, complaints received, and requested the consignee discard or destroy all affected devices. If the consignee further distributed the product, they were directed to forward a copy of the notice to those parties.
Initiated
2024-01-10
Root cause
Process control
Distribution
Worldwide distribution. US nationwide, Brazil, Canada, Switzerland, France, UK, Mexico, Netherlands, Saudi Arabia, Sweden, and South Africa
Recall number
Z-1057-2024
Recall status
Open, Classified
CDRH record ID
205565
FDA classified
2024-02-25
Recall event number
93821

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗