Covidien is issuing a voluntary field action for all Puritan Bennett 980 ventilators due to occasional GUI transient resets that last approximately 30 seconds.
Source facts
Action
Covidien sent an Urgent Field Safety Notice on January 12, 2015, to all affected customers. Customers were requested to forward the recall letter to anyone they may have transferred or assigned for use one of the potentially affected ventilators. Covidien is updating software that will modify ventilator alarms and status display messages during a GUI reset. The Puritan Bennett 980 Ventilator Operator Manual will be updated to delete,replace ventilator during a GUI transient reset, and to inform users to monitor the secondary status display until the primary display refreshes. Covidien will communicate with customers when the software and operator manual updates are available.
Customers with questions were instructed to contact Technical Support Department at 800-255-6774, option 4, and then option 1.
For questions regarding this recall call 303-876-8870.
Initiated
2015-01-12
Root cause
Software design
Terminated
2016-01-06
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada.
Recall number
Z-1058-2015
Recall status
Terminated
CDRH record ID
132617
FDA classified
2015-02-05
Recall event number
70239