A hardware update to replace the old PCB Assembly, Head Driver-G41 with PCB with overcurrent protection as a result of reports of the electric board inside the gantry showing signs of overheating, including thermal damage and smoke emission.
Source facts
Action
Siemens sent a Customer Advisory Notice Letter TH016/14/S, dated December 19, 2014, to end users that identified the product, problem, and actions to be taken. End users were advised to read the Field Safety Notice and include it with the User's Manual until the hardware update is completed.
Initiated
2014-12-19
Root cause
Component design/selection
Terminated
2017-01-26
Distribution
Worldwide Distribution - US Nationwide including Canada and Mexico.
Recall number
Z-1059-2015
Recall status
Terminated
CDRH record ID
132297
FDA classified
2015-02-05
Recall event number
70138