Z-1059-2019: Vacuum Unit for CATSmart Vacuum Pump. Fresenius CATSmart (Continuous Autotransfusion System) is indicate

Recall record Z-1059-2019 for Fenwal Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1059-2019
Company in FDA record
Fenwal Inc
FDA record creation date
Mar 26, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The optional Vacuum Unit may stop working and display the failure message "Failure vacuum unit" during use.

Source facts

Action
The firm, Fresenius Kabi, sent the initial distributor a notification on about 02/11/2019 via email. Consignees were notified on about 02/18/2019 via "IMPORTANT - VACUUM UNIT for CATSmart ISSUE" letter dated 2/7/2019. Instructions inducted that the device can still be used and there are not restrictions on the washing process. If an incident is experienced, customers are directed to contact Technical Service at 1-800-937-5060 and to provide the serial number of the device. Fresenius Kabi will be modifying the vacuum pump with a new cable mechanism to provide a secure long-term connection and updated firmware allowing for a consistent signal to the CATSmart device. The modifications will be implemented at customer locations and completed by September 1, 2019. Customers were also instructed to complete and return the Customer Reply form via fax to:1-847-550-0032. If you have any questions, please contact the Fresenius Kabi Technical Service at 800- 937-5060. prompt 2.
Initiated
2019-02-11
Root cause
Software design
Terminated
2020-06-23
Distribution
US Nationwide Distribution to states of: AZ, CA, FL, HI, ID, IL, IN, MD, MI, MN, MO, NC, ND, NE, NJ, NV, NY, OK, OR, PA, TN, TX, VA and PR; and Internationally to Canada.
Recall number
Z-1059-2019
Recall status
Terminated
CDRH record ID
171116
FDA record created
2019-03-26
Recall event number
82194

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗