Skip to content

Public FDA record

Z-1059-2022: VITEK 2 Systems and VITEK 2 with MYLA.

Recall record Z-1059-2022 for bioMerieux, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1059-2022
Company in FDA record
bioMerieux, Inc.
FDA record creation date
May 11, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software issue where results sent to the LIS via HL7 format for antibiotic screen tests and synergy tests do not include the user-corrected or AES-corrected interpretation. This can potentially lead to incorrect final screen/synergy test results at the LIS.

Source facts

Action
The recalling firm issued Field Safety Corrective Action (FSCA) #5615 dated 29-MAR-2022 to the impacted subsidiaries and distributors. It explained the affected product, issue, and informed the recipient the resolution for the issue has been implemented in VITEK 2 Systems software maintenance release 9MR3 launched 15-MAR-2022 and available in VILINK. It was recommended to convert HL7 users to VITEK 2 Software Version 9.03 and apply the software patch 9MR3 as quickly as possible. Workarounds were provided pending installation of 9MR3 to ensure there are no adverse impact to patients. The FSCA provided a risk assessment/health hazard assessment. The FSCA reported the VITEK 2 Systems software versions 8.01, 8.02, 9.01, 9.02, and 9.03 are available for distribution worldwide, however, only HL7 Protocol users may be impacted by the issue. All subsidiaries and distributors were listed in the FSCA. Their required actions included: (1) Immediately acknowledge receipt of this FSCA; (2) Identify all countries and customers for which they are responsible that are impacted by this FSCA (only HL7/BCI Connect users); (3) Update the enclosed Urgent Product Correction Notice and Customer Acknowledgment Form to include their local contact information and translate them, if necessary; (4) Distribute the Urgent Product Correction Notice and Customer Acknowledgement Form to all VITEK 2 HL7 customers; (5) Determine regulatory reporting requirements in accordance with local regulations and notify regulatory authorities as applicable; (6) The subsidiary/distributor is responsible to monitor action completion at customer sites and determine requirements for local closure of this FSCA in accordance with local regulations and procedures; (7) After all actions are complete, return the acknowledgement of completion for this FSCA. The due date for completion of the required actions and submitting the acknowledgment of completion is 29-JUN-2022. Additionally, Subsidiaries and Distribut
Initiated
2022-03-29
Root cause
Software design
Distribution
U.S. distribution of the FSCA to NC. OUS distribution of the FSCA to subsidiaries in Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Egypt, Finland, France, Germany, Greece, Hungary, India, Italy, Japan, Mexico, Netherlands, Norway, Philippines, Poland, Portugal, Russia, Serbia, Sweden, Singapore, South Africa, South Korea, Spain, Switzerland, Thailand, Turkey, and United Kingdom.
Recall number
Z-1059-2022
Recall status
Open, Classified
CDRH record ID
192933
FDA record created
2022-05-11
Recall event number
89929

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗