Open original FDA record ↗Recall record
- FDA identifier
- Z-1060-2013
- Company in FDA record
- Polymedco, Inc
- FDA classification date
- Apr 4, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The test strips in the affected product lots of the OC Light immunological fecal occult blood test are being recalled because they may not be stable through their labeled shelf-life. There may be a possibility of obtaining false negative results when using the affected lots.
Source facts
- Action
- Polymedco initiated phone calls to clients on 2/12/13. When the contact was made with the client, an URGENT DEVICE RECALL letter and Fax Form were sent by e-mail/fax. Mass mailing of the URGENT DEVICE RECALL letter dated February 12, 2013 and Fax Form was done by First Class USPS on 2/14/13-2/18/13. The recall affects the test strips only. For questions, please contact Polymedco Technical Support at 1-800-431-2123 ext 285.
- Initiated
- 2013-02-12
- Root cause
- Process control
- Terminated
- 2016-02-10
- Distribution
- Worldwide Distribution-USA (nationwide) including Puerto Rico, and the countries of Canada and Guam.
- Recall number
- Z-1060-2013
- Recall status
- Terminated
- CDRH record ID
- 116485
- FDA classified
- 2013-04-04
- Recall event number
- 64537