Skip to content

Public FDA record

Z-1060-2019: Integrated ECG cable with 3-lead leadwires labeled as the following: 1. ECG TRUNK CABLE, 3-LD W/ INTEG

Recall record Z-1060-2019 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1060-2019
Company in FDA record
GE Healthcare, LLC
FDA record creation date
Mar 26, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

The Integrated ECG cable with 3-lead leadwires can short circuit during defibrillation and conduct 25% of the defibrillation energy away from the patient.

Source facts

Action
Consignees were initially notified via telephone call on about 02/22/2019 and followed up with an "URGENT MEDICAL DEVICE CORECTION" letter. Instructions included to immediately stop use of the affected Integrated ECG cables and quarantine any on hand. GE Healthcare will replace all affected cables at no cost and a GE Healthcare representative will contact each customer to arrange for the collection of the affected cables and provide replacement of new cables. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171.
Initiated
2019-02-22
Root cause
Under Investigation by firm
Terminated
2020-08-19
Distribution
Worldwide Distribution - US Nationwide to AK, KS, MA, MA, ME, NJ, NJ, NM, NY, TX. Distributed worldwide to Chile, Finland, France, Germany, Italy, Korea (Republic Of), Russia, Spain, Sweden.
Recall number
Z-1060-2019
Recall status
Terminated
CDRH record ID
171255
FDA record created
2019-03-26
Recall event number
82272

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗