Open original FDA record ↗Recall record
- FDA identifier
- Z-1061-2013
- Company in FDA record
- Stryker Howmedica Osteonics Corp.
- FDA classification date
- Apr 5, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Stryker received reports from the field indicating that the handle detached from the cam of the Baseplate Impactor/Extractor Assembly.
Source facts
- Action
- Stryker sent Notification Letters/Product Accountablility Forms dated March 1, 2013 via FedEx to their Branches/Agencies on March 1, 2013.. Notification Letters/Product Recall Acknowledgement Forms dated March 4, 2013, were sent via FedEx to Hospital Risk Management on March 4, 2013. The Notification letters identified the product, the problem, and instructed customers on how to identify affected product; informs customers of the issues and potential hazards with the affected product and instructs them to contact their representative to arrange for return/replacement of any affected product. Customers are also asked to fax back the provided "Urgent Medical Device Removal Notification Acknowledgement Form" to Stryker within 5 days of receipt. Customers with questions were instructed to call 201-972-2100. For questions regarding this recall call 201-831-5523.
- Initiated
- 2013-02-23
- Root cause
- Nonconforming Material/Component
- Terminated
- 2015-06-15
- Distribution
- Worldwide Distribution - USA (nationwide) and internationally to Canada, Japan, Australia, China, New Zealand, France, Italy, Netherlands, Poland, Spain, Sweden, Switzerland and the United Kingdom.
- Recall number
- Z-1061-2013
- Recall status
- Terminated
- CDRH record ID
- 116723
- FDA classified
- 2013-04-05
- Recall event number
- 64655