Open original FDA record ↗Recall record
- FDA identifier
- Z-1062-2012
- Company in FDA record
- AGFA Corp.
- FDA classification date
- Feb 19, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The firm received a complaint that when using the DX-D Mobile DR X-Ray unit they noticed the detector cable was hot.
Source facts
- Action
- AGFA Corp. sent an "URGENT FIELD SAFETY NOTICE" letter dated January 16, 2012 to affected consignees. The letter described the affected product, concerns, issues and actions to be taken. Customers have been requested to place the Varian warning label included in a highly visible and suitable position on the cable of the portable detector. Customers were instructed to complete and return the feedback form by fax to 864-421-1664, as soon as possible. For questions about this matter contact your local AGFA HealthCare Technical Support at 877-777-2432.
- Initiated
- 2012-01-16
- Root cause
- Device Design
- Terminated
- 2012-05-10
- Distribution
- Nationwide Distribution (USA) including the states of: FL, IL, IN, KS, MI, MN, MS, NE,NY, OH, SC and WV.
- Recall number
- Z-1062-2012
- Recall status
- Terminated
- CDRH record ID
- 107178
- FDA classified
- 2012-02-19
- Recall event number
- 61052