Open original FDA record ↗Recall record
- FDA identifier
- Z-1063-2015
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Feb 7, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a low probability the scanner arm will become completely detached from the scanner column.
Source facts
- Action
- Siemens sent a Customer Safety Advisory Notification to all affected sites on January 12, 2015, certified mail, with return receipt. The letter identified the affected product, the issue, and potential risk. Customers were asked to immediately discontinue use of the affected product. A Customer Service Engineer will contact customers to schedule an appointment to make the necessary repairs. Questions should be directed to a local service support person for information regarding timelines and status. Customers are to share the information with all personnel, who need to be aware of the issue.
- Initiated
- 2015-01-12
- Root cause
- Vendor change control
- Terminated
- 2015-06-10
- Distribution
- Worldwide Distribution -- US, including the states of TX and MN, and the territory of P.R.; and, the countries of China, Germany, Russian Fed, Saudi Arabia, Taiwan, France, Thailand, and Italy.
- Recall number
- Z-1063-2015
- Recall status
- Terminated
- CDRH record ID
- 133279
- FDA classified
- 2015-02-07
- Recall event number
- 70361