Open original FDA record ↗Recall record
- FDA identifier
- Z-1064-2011
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Feb 8, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Potential for Mistreatment-- when treatment plans intended for non-ARTISTE machinese are loaded, if the plans contain unspecified beam parameters, attributes or unknown block codes, the therapy system may remove those parameters.
Source facts
- Action
- The firm,Siemens, sent two Customer Advisory letters (TH004/09/S and TH008/09/S) contained in the Update Instructions to all consignees/customers with active devices. The letters describe the product, problem and action to be taken by the customers. The customers were instructed to follow the updated instructions. Siemens service representative will sign a document noted as CORRECTED upon completion of the software installation. If you have any questions, call 1-925-602-8157.
- Initiated
- 2009-07-30
- Root cause
- Software design
- Terminated
- 2011-04-12
- Distribution
- Worldwide distribution: USA and countries including: Germany, France, Poland, Norway, Australia, Malaysia, Spain, New Zealand, the UK, India, Republic of Korea, Canada, Ireland, Italy, Egypt, Belgium, Philippines, Saudi Arabia, Netherlands, Hungary, Norway and South Africa.
- Recall number
- Z-1064-2011
- Recall status
- Terminated
- CDRH record ID
- 84556
- FDA classified
- 2011-02-08
- Recall event number
- 53054